Labratory Incubator chamber volume 404 litres- Evo Line
0.00SAR
FRIOCELL 404 - EVO Line
Product Details
- precise, adaptable to nature, economical - laboratory incubator
- chamber volume 404 litres
- temperature range: 0°C to 100°C
- patented cooling system (ecological working medium R134a) offers accurate and fast simulation of natural processes and reduces the risk of evaporation of samples, the devices may be completed with various types of inner chamber lighting
- high operation comfort, precise temperature control and the possibility of achieving extremely short times of temperature changes as well as short times of temperature equalization in the chamber after the door opening; standard equipment also includes inner glass door and possibility of fan speed regulation
- suitable for demanding and accurate specimen culture testing and processes
- wide selection of optional equipment and accessories
- output for connecting a printer or PC
- possibility of validation (IQ, OQ)
- precise, adaptable to nature, economical - laboratory incubator
- chamber volume 404 litres
- temperature range: 0°C to 100°C
- patented cooling system (ecological working medium R134a) offers accurate and fast simulation of natural processes and reduces the risk of evaporation of samples, the devices may be completed with various types of inner chamber lighting
- high operation comfort, precise temperature control and the possibility of achieving extremely short times of temperature changes as well as short times of temperature equalization in the chamber after the door opening; standard equipment also includes inner glass door and possibility of fan speed regulation
- suitable for demanding and accurate specimen culture testing and processes
- wide selection of optional equipment and accessories
- output for connecting a printer or PC
- possibility of validation (IQ, OQ)
Inner dimensions - Chamber, stainless steel
volume 404 l depth 530 mm width 540 mm height 1415 mm
volume | 404 l |
depth | 530 mm |
width | 540 mm |
height | 1415 mm |
Volume of chamber space
530 l (aprox.)
530 l (aprox.) |
Trays, stainless steel *)
Min. distance between trays 70 mm number 19/2 max./usual Storage area (w × d) 520 x 485 mm
Min. distance between trays | 70 mm |
number | 19/2 max./usual |
Storage area (w × d) | 520 x 485 mm |
Number of outer metal doors
number 1 pcs
number | 1 pcs |
Number of inner glass doors
number 1 pcs
number | 1 pcs |
Admissible weight of trays
together inside the oven 100 kg per 1 tray 30 kg
together inside the oven | 100 kg |
per 1 tray | 30 kg |
Electricity data
Max. power 1700 W IP 20 mains 50/60Hz 230 V
Max. power | 1700 W |
IP 20 | |
mains 50/60Hz | 230 V |
Temperature data
Cooling/down time from 22 °C to 10 °C 19 min Heating/up time to 37 °C from the ambient temperature 13 min Working temperature from 0,0 to 100 °C Temperature accuracy - in time 0,3 ± °C Temperature accuracy - in space at 10 °C 1 ± °C
Cooling/down time from 22 °C to 10 °C | 19 min |
Heating/up time to 37 °C from the ambient temperature | 13 min |
Working temperature from 0,0 to | 100 °C |
Temperature accuracy - in time | 0,3 ± °C |
Temperature accuracy - in space at 10 °C | 1 ± °C |
Recovery time after 30 s door open
at 37 °C 2 min at 50 °C 4 min
at 37 °C | 2 min |
at 50 °C | 4 min |
Heat emission
at 37 °C 123 W
at 37 °C | 123 W |
Outer dimensions, incl. door and handle, Feet (N) or Rollers (K)
depth 885 mm width 1100 mm height 1890 (K) mm
depth | 885 mm |
width | 1100 mm |
height | 1890 (K) mm |
Weight
brut 390 kg nett 230 kg
*) Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber.Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.
Changes in the design and make reserved.
The values may differ depending on specific charge and media parameters.
brut | 390 kg |
nett | 230 kg |
*) Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber.Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.
Changes in the design and make reserved.The values may differ depending on specific charge and media parameters.