- chamber volume 111 litres
- temperature range: 0°C to 100°C
- patented cooling system (ecological working medium R134a) offers accurate and fast simulation of natural processes and reduces the risk of evaporation of samples, the devices may be completed with various types of inner chamber lighting
- high operation comfort, precise temperature control and the possibility of achieving extremely short times of temperature changes as well as short times of temperature equalization in the chamber after the door opening; standard equipment also includes inner glass door and possibility of fan speed regulation
- suitable for demanding and accurate specimen culture testing and processes
- wide selection of optional equipment and accessories
- output for connecting a printer or PC
- possibility of validation (IQ, OQ)
- precise, adaptable to nature, economical - laboratory incubator
- chamber volume 111 litres
- temperature range: 0°C to 100°C
- patented cooling system (ecological working medium R134a) offers accurate and fast simulation of natural processes and reduces the risk of evaporation of samples, the devices may be completed with various types of inner chamber lighting
- high operation comfort, precise temperature control and the possibility of achieving extremely short times of temperature changes as well as short times of temperature equalization in the chamber after the door opening; standard equipment also includes inner glass door and possibility of fan speed regulation
- suitable for demanding and accurate specimen culture testing and processes
- wide selection of optional equipment and accessories
- output for connecting a printer or PC
- possibility of validation (IQ, OQ)
Inner dimensions - Chamber, stainless steel
volume 111 l depth 370 mm width 540 mm height 530 mm
volume | 111 l |
depth | 370 mm |
width | 540 mm |
height | 530 mm |
Volume of chamber space
163 l (aprox.)
163 l (aprox.) |
Trays, stainless steel *)
Min. distance between trays 70 mm number 7/2 max./usual Storage area (w × d) 520 × 335 mm
Min. distance between trays | 70 mm |
number | 7/2 max./usual |
Storage area (w × d) | 520 × 335 mm |
Number of outer metal doors
number 1 pcs
number | 1 pcs |
Number of inner glass doors
number 1 pcs
number | 1 pcs |
Admissible weight of trays
together inside the oven 50 kg per 1 tray 20 kg
together inside the oven | 50 kg |
per 1 tray | 20 kg |
Electricity data
Max. power 700 W IP 20 mains 50/60Hz 230 V
Max. power | 700 W |
IP 20 | |
mains 50/60Hz | 230 V |
Temperature data
Cooling/down time from 22 °C to 10 °C < 21 min Heating/up time to 37 °C from the ambient temperature 11 min Working temperature from 0,0 to 100 °C Temperature accuracy - in time 0,2 ± °C Temperature accuracy - in space at 10 °C 0,5 ± °C
Cooling/down time from 22 °C to 10 °C | < 21 min |
Heating/up time to 37 °C from the ambient temperature | 11 min |
Working temperature from 0,0 to | 100 °C |
Temperature accuracy - in time | 0,2 ± °C |
Temperature accuracy - in space at 10 °C | 0,5 ± °C |
Recovery time after 30 s door open
at 37 °C 5 min at 50 °C 6 min
at 37 °C | 5 min |
at 50 °C | 6 min |
Heat emission
at 37 °C 70 W
at 37 °C | 70 W |
Outer dimensions, incl. door and handle, Feet (N) or Rollers (K)
depth 750 mm width 760 mm height 1045 (N) mm
depth | 750 mm |
width | 760 mm |
height | 1045 (N) mm |
Weight
brut 136 kg nett 106 kg
*) Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber.Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.
Changes in the design and make reserved.
The values may differ depending on specific charge and media parameters.
brut | 136 kg |
nett | 106 kg |
*) Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber.Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.
Changes in the design and make reserved.The values may differ depending on specific charge and media parameters.