laboratory incubator - always creates the right atmosphere
chamber volume 50 litres
temperature range: from 5°C above ambient temperature to 60°C
uncontrolled relative humidity: up to 90 ± 5% RH at 37°C
CO2 concentration: 0.2% to 20%
CO2 sensor: no-deviation infrared sensor (IR)
wide selection of optional equipment and accessories
output for connecting a printer or PC
possibility of validation (IQ, OQ)
Inner dimensions
volume
50 l
depth
300 mm
width
402 mm
height
402 mm
Shelves (trays)*
standard equipment
3 pcs
Number
3 max. pcs
External dimensions
depth
460 mm
width
534 mm
height
717 mm
Weight
brut
50 kg
nett
38 kg
Temperature data
60 °C
Temperature deviations at 37 °C max.
in time
± 0,1 °C
in time
± 0,1 °C
in space
± 0,25 °C
CO2 Control
0,2 - 20 % CO2
Control accuracy
in space
± 0,25 %
measurement
± 0,1 %
measurement
± 0,1 %
Recovery rate
<=0,8 % / min
Required CO2 pressure
0,3-0,7 bar / psi
Relative humidity
humidity dish volume
0,5 l
Humidity level at 37 °C, 5 °C CO2
standard
95 % RH
Electric parameters
max. output
0,35 kW
Voltage (50/60 Hz)
230 V
consumption at 37 °C
7,2 W
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off. Changes in the design and make reserved. The values may differ depending on specific charge and media parameters.
Items have been added to cart.
One or more items could not be added to cart due to certain restrictions.