





INCUCELL V 111 - ECO line
0.00SAR
INCUCELL V 111 - ECO line
Product Details
- natural, accurate - laboratory incubator
- chamber volume 111 litres
- temperature range from 10°C above ambient to 100°C
- characterized by noiseless operation and fine air flow in the usable space
- high operation comfort, precise temperature control and short chamber temperature equalization times after the door opening
- high-quality and safe drying, heating and tempering of common laboratory materials in laboratory and industrial areas, suitable for safe processing of microbiological cultures
- standard version with microprocessor control system Fuzzy logic, controls placed on a foil keyboard, process information displayed on LED display
- wide selection of optional equipment and accessories
- output for connecting a printer or PC
- possibility of validation (IQ, OQ)
Interior of stainless steel material No. 1.4301
volume 111 l width 540 mm height 530 mm Tray
standard equipment 2 pcs storage area (w × d) 520 × 335 mm racks 7 max. number distance between guide rails 70 mm Maximal weight of the load *)
inside the oven 50 kg/oven per tray 20 kg/tray Door
1 pcs External dimensions (including door and handle)
depth 680 mm diameter of the air branch outer/inner 52/49 mm width 760 mm 860 (N) Package dimensions (three layers carton)
depth 750 mm width 830 mm depth 1060 mm Weight
brut 87 kg nett 75 kg Working temperature (beginning of the regulation)
from 10 °C above ambient temp. to: 100 °C Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door
time variation cca (±) <=0,2 °C space deviation cca (±) <=0,3 °C Time required to reach 37 °C with closed air flap and voltage 230 V
51 min Heat emission at 37 °C
95 W Air exchange speed at 37 °C
49 per hour Electricity - mains 50/60 Hz
max. power 0,7 kW nominal voltage 230 V current 3 A power by stand by 5 W *) Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber. Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.
Changes in the design and make reserved.
The values may differ depending on specific charge and media parameters.