INCUCELL 222 - ECO line
0.00SAR
INCUCELL 222 - ECO line
Product Details
- natural, accurate - laboratory incubator
- chamber volume 222 litres
- temperature range from 5°C above ambient to 100°C
- characterized by noiseless operation and fine air flow in the usable space
- high operation comfort, precise temperature control and short chamber temperature equalization times after the door opening
- high-quality and safe drying, heating and tempering of common laboratory materials in laboratory and industrial areas, suitable for safe processing of microbiological cultures
- standard version with microprocessor control system Fuzzy logic, controls placed on a foil keyboard, process information displayed on LED display
- wide selection of optional equipment and accessories
- output for connecting a printer or PC
- possibility of validation (IQ, OQ)
Interior of stainless steel material No. 1.4301
volume 222 l depth 520 mm width 540 mm height 760 mm Tray
standard equipment 2 pcs storage area (w × d) 520 × 485 mm racks 10 max. number distance between guide rails 70 mm Maximal weight of the load*)
inside the oven 70 kg/oven per tray 30 kg/tray Door
1 pcs External dimensions (including door and handle)
depth 830 mm diameter of the air branch outer / inner 52/49 mm width 760 mm 1090 (N) Package dimensions (three layers carton)
depth 890 mm width 820 mm depth 1260 mm Weight
brut 116 kg nett 100 kg Working temperature (beginning of the regulation)
from 5 °C above ambient temp. to: 100 °C Temperature accuracy according to DIN 12 880 T2 at working temperature with closed air flap and door
time variation cca (±) <=0,2 °C space deviation cca (±) < 1 °C Time required to reach 37 °C with closed air flap and voltage 230 V
79 min Heat emission at 37 °C
45 W Air exchange speed at 37 °C
5 per hour Electricity - mains 50/60 Hz
max. power 0,5 kW nominal voltage 230 V current 2 A power by stand by 5 W *) Approx. 50 % of the tray area can be filled in a way a uniform air circulation is enabled inside the chamber. Note: All technical data are related to 22 °C ambient temperature and +-10 % voltage swing (if not specified).
Changes in the design and make reserved.