DUROCELL 55 - ECO line
0.00SAR
Product Details
- highly resistant laboratory drying oven
- chamber volume 55 litres
- temperature range from 5°C above ambient to 125°C
- natural fine airflow in the work area, noiseless operation
- high-quality and safe drying, heating and tempering of laboratory materials in the laboratory and industrial areas
- high operation comfort, precise temperature control and short chamber temperature equalization times after the door opening
- wide selection of optional equipment and accessories
- output for connecting a printer or PC
- possibility of validation (IQ, OQ)
Interior of stainless steel material No. 1.4301
volume | 55 l |
depth | 370 mm |
width | 400 mm |
height | 350 mm |
Trays
standard equipment | 2 pcs |
storage area (w × d) | 380×335 mm |
racks | 4 max. no |
distance between guide rails | 70 mm |
Maximal weight of the load*)
inside the oven | 50 kg/oven |
per tray | 20 kg/tray |
Door
1 pcs |
External dimensions (including door and handle)
depth | 680 mm |
diameter of the air branch outer/inner | 52/49 mm |
width | 620 mm |
height (inclusive foots - N) | 680 (N) mm |
Package dimensions (three layers carton)
depth | 760 mm |
width | 700 mm |
depth | 880 mm |
Weight
brut | 66 kg |
nett | 55 kg |
Working temperature (beginning of the regulation)
from 5 °C above ambient temp. to: | 125 °C |
Temperature accuracy according to DIN 12 880 T2, at working temperature with closed air flap and door
time variation cca (±) | 0,3 °C |
space deviation cca (±) % of the reached temperature | 2 % |
Time required to reach 100 °C with closed air flap and voltage 230 V
41 min |
Heat emission at 100 °C
380 W |
Air exchange speed at 100 °C
8 per hour |
Electricity – mains 50/60 Hz
max. power input | 1,2 kW |
nominal voltage | 230 V |
current | 5,2 A |
input in stand by mode | 5 W |
Note: All technical data are related to 22 °C of ambient temperature and ±10 % voltage swing rate ventilator 100 %, defrost off, lighting off.
Changes in the design and make reserved.The values may differ depending on specific charge and media parameters.